![]() Conclusions: The epoc system is fit for use in the surgical and ICU setting for the measurement of all analytes except for creatinine. Creatinine showed the largest bias relative to laboratory analyzers, Abbott Architect c8000 Jaffe method (13.54%, 95% CI = 5.43, 21.65) and Roche Cobas c702 enzymatic method (30.01%, 95% CI = 12.64, 47.38). Because creatinine is a nonprotein end-product of creatine phosphate, which is used in skeletal muscle contraction, the daily production of creatine, and the following product, creatinine, depends on muscle mass, which fluctuates very little. Dictionary com s List of Every Word of the Year. When muscles use this energy, muscle tissue breaks down, releasing creatinine (a toxin) into the blood. Your body makes and uses creatine, a chemical, to provide energy to muscles. Hematocrit showed a bias of -6.76% (95% CI = -8.91, - 4.61) compared to the HemataSTAT-II method, whereas calculated total hemoglobin showed a bias of 1.51% (95% CI = -1.04, 4.06) against the Sysmex XN-10 hematology analyzer. The creatinine test is used to measure the amount of creatinine in the blood. Creatinine Testing Handheld istat creatinine system. A serum creatinine blood test measures levels of creatinine, a waste product in your blood. Results: Mean bias was <5% for blood gases, electrolytes, lactate and glucose. Method comparison was performed for all analytes on the epoc System against the RAPIDPoint 500 Blood gas analyzer or laboratory analyzers where applicable. ![]() ![]() Venous or arterial whole blood samples collected in balanced heparinized syringes were obtained from 69 patients (35 females, 34 males) predominantly (77%) from the surgical unit and intensive care unit (ICU).